Product Detail :
|
Supply of Srph154052-Single Chamber Rate Responsive Pacemaker (System Should Be Mri Conditional) With Appropriate Bipolar Mri Conditional Lead. Should Have The Latest Technological Features To Provide Proportionate, Smooth And Appropriate Response Of Pacing Rate To Increased Metabolic Demand. Should Be Suitable For Use Inadult And Children. Should Have Sufficient Programming Flexibility For Using In Atrium (Where Sensitivity Should Be Adjustable To 0.25Mv Or Lower) And / Or Ventricle And Should Be Supplied With Low Profile Atrial / Ventricular Mri Compatible Lead. (To Be Specified When Taking Delivery). The Lead Should Be Low Profile (7F Or Smaller Introducer) And Should Be Steroid Eluting. Should Be Capable Of High Energy Output Pacing With Pulse Amplitude Of Up To At Least 6.0V And Pulse Width Up To Atleast 1.0Ms. The Pacing System Should Have Maximum Flexibility Of Programming, Data Storage And Display For Ease Of Clinical Follow Up And Diagnosis Of Malfunction. Should Be Covered By Life Warranty. (Providing Pacing Therapy For The Duration Of Patients Life Time). Should Provide Free Of Charge The Latest Programmer With Appropriate Software Which Should Have Facilities For Simultaneous Display Of Multiple Surface Ec
|
Tender Detail : |
Single Chamber Rate Responsive Pacemaker |